logo

products details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Flammability Testing Equipment
Created with Pixso.

0.1MPa Flammability Testing Equipment

0.1MPa Flammability Testing Equipment

Brand Name: HeJin
Model Number: HVF-3
MOQ: 1 Set
Price: Negotiatable
Payment Terms: T/T, L/C
Supply Ability: 15 set per month
Detail Information
Place of Origin:
China
Certification:
Calibration Certificate(cost additional)
Flame Power:
50W&500W
Temperature Testing:
Max. 1100℃
Combustion Angle:
0°,20°,45°
Standards:
UL94, IEC60695-11-4, IEC60695-11-3
Working Voltage:
220V/50Hz
Gas Pressure:
0.1MPa
Packaging Details:
Plywood
Highlight:

0.1MPa Flammability Testing Equipment

,

0.1MPa Flammability Testing

,

Flammability Testing Equipment 0.1MPa

Product Description

UL94 Flammability Testing Equipment Horizontal And Vertical Burning With 50w And 500w Flames

 

Standard:

UL94 ‘Tests for Flammability of Plastic Materials for Parts in Devices and Appliances’, IEC60695-11-3 ‘Fire hazard testing – Part 11-3: Test flames – 500 W flames – Apparatus and confirmational test Methods’, IEC60695-11-4 ‘Fire hazard testing – Part 11-4: Test flames – 50 W flames – Apparatus and confirmational test Methods’.

 

Application:

It simulates the influence of early-stage flame when there is a fire around the electric and electronic products, so that to judge the igniting danger degree. It is mainly used in plastic and other non-metallic material sample, solid material. It is also applicable in the Horizontal, vertical flammability test of the relative combustion characteristic of foam plastics whose density is no-less than 250kg/m according to ISO845 test method.

 

Parameters:

Parameters

Model

HVF-1 HVF-2 HVF-3(Touch screen)
Working voltage 220V/50Hz 220V/50Hz 220V/50Hz
Power/grade 50W or 500W 50W and 500W double power switching 50W and 500W double power switching
Control operation mode Electrical control, button operation Electrical control, button operation PLC control, 7 inch color touch screen operation, with infrared remote control
Blast burner standard

IEC60695-11-3/IEC60695-11-4/10,UL94,

inner diameter:Φ9.5±0.3mm,length:100mm, made of brass

Gas pressure 0.1MPa
Gas flow

50W:105±5ml/min or 18±1mm;

500W: 965±30ml/min or 117±3 mm, flow meter adjustment

U type pressure gauge back pressure water column height 50W:10mm; 500W:125±5 mm
Test angle 50W: 0°,45°; 500W: 20°
Burning time 0-9999X0.1S Can be preset
Burning times 0-9999 times, Can be preset
Residual flame/Residual glow 0-9999X0.1S Can be displayed
Test flame calibration Temperature measuring copper head: 50W: Ф5.5mm±0.01mm, weight before drilling 1.76 ± 0 .01 g; 500W: Ф9mm±0.01mm, weight before drilling 10 ± 0 .05g
Thermocouple: The American original Omega φ0.5 armour nickel-chromium/nickel aluminum wire K-Thermocouple, temperature resistance 1100°
Distance between the bunsen burner port to temperature measuring copper block: 50W: 10±1 mm; 500W: 55±1 mm

Flame height: 50W: 20±2mm(there is no internal and external flame); 500W: Inner flame:40±2mm, Outer flame:125±10mm

Temperature rise time from100±2°to 700°±3°50W: 44±2S; 500W: 54±2S
Lighting Equip with lighting in the test chamber, with control
Test chamber and dimension > 0.75 cube, black inwall, with exhaust fan, dimension:1100*550*1200mm, 140kg
 
0.1MPa Flammability Testing Equipment 0
Ratings & Review

Overall Rating

5.0
Based on 50 reviews for this supplier

Rating Snapshot

The following is the distribution of all ratings
5 stars
0%
4 stars
0%
3 stars
0%
2 stars
0%
1 stars
0%

All Reviews

D
Detecting Leakage Device for Precise Leak Detection in Industrial Environments
United States Dec 1.2025
This helium leak test machine works far better than the previous system we used. The accuracy is outstanding, and our testing efficiency improved by over 30%.
D
Detecting Leakage Device for Precise Leak Detection in Industrial Environments
Germany Oct 22.2025
Die Verarbeitungsqualität ist hervorragend. Wir haben zwei Geräte für Membranspeichertests gekauft, und beide arbeiten mit bemerkenswerter Stabilität und Wiederholgenauigkeit.